Product Details

  salt spray test is against the surface of a variety of materials by plating, anodizing, painting, anti-rust oil anti-corrosion treatment, test the corrosion resistance of their products。

  the implementation of standards and test methods

  GB / T 2423.17-1993 salt spray test

  GB / T 2423。18-2000 salt spray test

山东11选5开奖结果   GB / T 10125-1997 salt spray test

  ASTM.B117-97 salt spray test

  JIS H8502 salt spray test

山东11选5开奖结果   IEC68-2-11 salt spray test

  IEC68-2-52 1996 salt spray test

  GB。10587-89 salt spray test

  CNS.4158 salt spray test

  CNS.4159 CASS copper accelerated acetic acid salt spray test

  GB / T 12967.3-91 CASS copper accelerated acetic acid salt spray test

  Specifications

  Model WS-60B WS-90B WS-120B WS-160B WS-200B

  test chamber box size (cm) 60x45x40 90x60x50 120x80x50 160x100x50 200x120x60

  Test outdoor box size (cm) 107x60x118 141x88x128 190x110x140 230x130x140 270x150x150

  test chamber temperature saline test method (NSS ACSS) 35 ℃ ± 1 ℃ / corrosion test method (CASS) 50 ℃ ± 1 ℃

  pressure tank temperature saline test method (NSS ACSS) 47 ℃ ± 1 ℃ / corrosion test method (CASS) 63 ℃ ± 1 ℃

山东11选5开奖结果   brine temperature of 35 ℃ ± 1 ℃ 50 ℃ ± 1 ℃

  Laboratory capacity 108L 270L 480L 800L 1440L

  brine tank capacity 15L 25L 40L 40L 40L

  the concentration of sodium chloride salt concentration of 5% or 5% concentration of sodium chloride solution was added 0.26 grams per liter of copper chloride (CuCl2 2H2O)

  Air Pressure 1。00 ± 0。01kgf / cm2

山东11选5开奖结果   spray amount 1.0 ~ 2.0ml / 80cm2 / h (collected at least 16 hours, the average value)

  laboratory relative humidity is above 85%

  pH PH 6。5 ~ 7。2 3。0 ~ 3。2

  spray mode programmable spray (including continuous and intermittent spray spray)

  Power AC220V1Φ10A AC220V1Φ15A AC220V1Φ20A AC220V1Φ20A AC220V1Φ30A

  Temperature range ambient + 5 ℃ ~ 65 ℃ temperature uniformity ≤ ± 2 ℃

  Temperature fluctuation ≤ ± 0。5 ℃ Salt spray settlement 1 ~ 2ml / 80cm2。h

  spray in a continuous cycle optionally test the timing 1 ~ 9999 (S, M, H) Adjustable

  temperature control Omron or Japan Chemical LED digital display P, I, D + S, S, R. Microcomputer integrated controller tower spray system plus no crystal nozzle spray device (mist particles finer and uniform distribution)

  platinum resistance temperature sensor .PT100Ω / MV heating system independent systems, corrosion-resistant titanium electric heating type heater

  1 ~ 9990H spray time and cycle temperature and humidity inside with adjustable temperature and humidity mercury thermometer

  salt was collected and standard measurement standard funnel cylinder saline salt solution preheat temperature and inside temperature equilibrium

  inside and outside the box material high strength, high temperature, acid, advanced PVC plate cover material imported corrosion-resistant, high strength transparent acrylic plate

山东11选5开奖结果   security leakage, short circuit, over-temperature, water, end of the trial, over-current protection standard configuration rod, V-type sample rack 1 pay, the nozzle 2, funnel, measuring cylinder 2 sets

  other accessories are stainless steel, brass or corrosion, high temperature resistant materials supply voltage 220V 50HZ

  ambient temperature "5 ℃ ~ + 30 ℃ ≤85% RH sealed with water seal structure, no salt overflow

  5 Features

  natural environmental exposure test artificial accelerated salt spray environmental testing; artificial salt spray test is the use of having a certain volume of space of test equipment - salt spray chamber, the volume of space within which use artificial methods, resulting in salt spray environment of salt spray corrosion resistance aging quality of the product。 Compared with the natural environment, the chloride salt spray environment, which can be several times or several times the general content of the natural environment of salt spray, corrosion rate greatly improved the product salt spray test, the results The time is also greatly reduced。

  product conforms: CNS, ASTM, JIS, ISO standard salt spray test is against the surface of a variety of materials by coating, plating, anodizing, anti-rust oil and anti-corrosion treatment, test the corrosion resistance of their products.

  imported intelligent temperature control instrumentation。 The meter high intelligence, multiple and fuzzy PID control operation, fast setting, to achieve a smoother control output and higher control accuracy。

  with automatic timing devices, test laboratory after reaching a predetermined temperature, automatic spray, Time timer automatically stops。

  laboratory may also be transparent cover, always monitor the spraying status and working conditions of the sample.

  the use of non-crystalline spray nozzle mouth, no danger of clogging of salt.

  There are dual pressure adjustment and over-temperature protection device.

  the structure and characteristics 1, inside and outside the gall of the studio equipment are used epoxy fiberglass integrally molded, light weight, corrosion resistance, high strength, no leakage, the surface smooth, smooth and easy to clean and so on.

  2, a water seal in the form of salt spray will not spill duct heating jacket, make studio warming up fast, temperature distribution。

  3, salt box cover is designed to 110 ~ 112 ℃ angle, not to the cold tank condensation dripping water directly on the sample.

  4, spray method is a spray tower, a spray tower placed on the bottom surface of the middle of the studio。 Nozzle material is quartz glass baked material fine mist sprayed salt spray natural subsidence was evenly distributed。

  5, control system equipment is placed in the top right box, user-friendly operation, set the power switch, ranked fog switch, spray switch on the front panel, two self-tuning PID digital instrumentation, and equipment operation time timer and the gap spray timer circuit board in the cabinet also ad hoc over-temperature protection.

  6 Test principle

  corrosive solution is sprayed into the compressed air fog, mist which surrounded the test sample in a test chamber.

  test can be continuous or cycle.

  corrosion resistance of each sample time is defined as the sample appears to corrosion, longer show the better corrosion resistance.

  salt spray corrosion test chamber Technical data

  7 Practice

  1. salt spray me, please turn on the power supply and air compressor, air hose connection.

  2. The pure water or distilled water is added behind cabinets and cabinet inlet, until the low water level light on the panel off date, or can not operate normally.

  3. drain and an exhaust pipe connected, as shown on the previous indicators is on the left.

  4. The isolation tank with water to the plate position, in order to avoid salt spray leakage.

  5. Test solution deployment modulation methods:

  A. overburden, neutral salt spray test (NSS test)

  a。 Saline solution using sodium chloride (chemically pure, analytical grade) and distilled or deionized water at a concentration of (5% ± 0。1)% (mass percentage)。 Atomized liquid collected, in addition to the bezel turned back part, can not be reused;

  b. PH value of salt solution before atomization at 6.5 ~ 7.2 (35 ± 2 ℃) between. The preparation of a salt solution, can be chemically pure solution of dilute hydrochloric acid or sodium hydroxide to adjust the PH value, but still comply with the concentration of a point.

  B. metal cover layer, copper accelerated acetic acid salt spray test (CASS test)

  a. Sodium chloride was dissolved in distilled or deionized water at a concentration of 50 ± 5g / L.

山东11选5开奖结果   b。 The added to a solution of copper (CuCl2 · 2H2O) chloride at a concentration of 0。26 ± 0。02g / L (i。e。, 0。205 ± 0。015g / L anhydrous cupric chloride)。

  c. Adding an appropriate amount in a solution of acetic acid salt spray test chamber to ensure the pH value of the liquid collected 3.1 to 3.3. As before spraying the solution pH is 3.0 to 3.1, the pH of the collected solution is generally in the range of 3.1 to 3.3. PH meter with a pH value of the solution was measured, it can also be proofread by the acidity of the read 0.1pH value change of precision pH test paper as a routine test. PH of the solution can be used to adjust the glacial acetic acid or sodium hydroxide.

  d. In order to avoid nozzle clogging, the solution must be filtered before use.

  6. Pour salt brine liquid supplement bottle, automatically filling brine into the laboratory spray tower, so that the drug flow to the salt spray tower.

  7. The wet ball cup water, covered with gauze wet bulb thermometer, wet gauze ball ends cup placed.

  8。 Place the specimen or sample on the shelf: display the desired angle in accordance with standard furnishings, such as the standard specimen 130 × 70 (mm) Available 15 degrees, 30 degrees inclined。

  9. Setting the test temperature: according to the desired standard setting (button "▲" to increase, press "▼" to decrease)

  a。NSS, AASS test: the test chamber temperature 35 ℃; saturated air barrel temperature of 37 ℃ (35 ℃ ~ 40 ℃)。

  b.CASS test: the test chamber temperature 50 ℃; saturated air barrel temperature of 55 ℃ (50 ℃ ~ 55 ℃).

山东11选5开奖结果   10。 Setting the test time: 0 ~ 999Hr (H: the time; the key "+" to increase; press "-" to decrease)

  11. Press the power button operation, pre-heating in advance, to reach the set temperature. Note: The need to test and handled with care so as not to damage the cover lid.

  12。 Press the spray button:

山东11选5开奖结果   a, will be out in front of the open compressor valve, the pressure was adjusted to 0.2 ~ 0.4MPa primary pressure;

  b, the regulator raised pressure of 0.07 ~ 0.17MPa, the pressure gauge can be learned (clockwise to increase, counterclockwise to decrease).

  13. Press the timer button, depending on the timing of the set time.

  14. After the test, according to the order of each switch is turned off.

  15。 If the test anomalies, refer judge dysfunction treatment table。

  8 Test preparation

  1, add a little heat sink inside the bottom of the studio distilled or deionized water to prevent chapped aging of the box when heated.

  2, the water seal groove upper box four weeks should be added the right amount of distilled or deionized water, not too full, not too little, to close the lid, the water and salt spray is better not spill.

  3, the air saturator (stainless steel barrels) of added distilled or deionized water before adding water to open (Figure 1) plus the purge valve and saturator valve until the water level for the water glass box 4 / 5 position is appropriate, when added to a predetermined level, it must close the bleed valve while closing the saturator plus valve。 Due to long-term testing, saturator water evaporation consumption when lowering the water level to 2/5 should timely pay, to prevent water shortages lead to heating pipes empty roasting, baking bad saturator and internal plus hot components。

  4, check the water storage tanks and the working chamber is connected between the sprayer intact, the good preparation of a salt solution (typically formulated at 5% concentration) was added to the storage tank.

  5, check the box rear exhaust pipe exhaust fog fog conditions; trachea is off, whether at the outlet pass to the sewer; pipeline is blocked; so as not to affect the salt mist emissions.

  6, stand a good connection pipe inside the funnel casing, check and set the funnel between the fog is smooth intact, it does not affect the collection of salt spray settlement.

山东11选5开奖结果   7, check the gas source and saturated the connection pipe; gas source connecting pipe with nebulizer is off, to prevent spills or gas supply shortage in the state。

  8, open the lid, the sample is properly placed between the working chamber of the sample holder (away from the surrounding walls 10%), to maintain a certain distance from sample to sample, the sample is exposed surface area & gt; 98%, and then close the cover。

  9 Maintenance and Other Considerations

山东11选5开奖结果   1, after the end of each test, you should open oil-water separator switch, the oil and the water put the net, and wipe clean。

  2, if a long-term test shall not saturated the water put the net, during normal use, should periodically replace the water in the saturator。

  3, when the long-term need, if re-do the test, all electrical and respond to water and gas supply systems to make a specific system check, failure to prevent the boot after.

  4, found that the nozzle clogging dirt, promptly remove the nozzle, with alcohol, xylene, or 1: 1 hydrochloric acid wash or superfine wire dredge careful not damage the wall of the nozzle opening, so as not to affect the spray quality.

山东11选5开奖结果   the implementation of standards

  The product can meet GB9789-88 "metal corrosion test under organic and other non-metallic covering layer is generally condensation conditions" test requirements。

  11 pressure regulator

山东11选5开奖结果   a: inlet pressure regulator from the first stage regulator manually, the general control 2 ~ 4kg / cm2

  b:。 Spray pressure by a reducing valve adjusted to 0。5 ~ 1。7kg / cm2, the classification has been adjusted adjusted at the factory, users generally will not need to adjust

  solution preparation program

  solution preparation salt spray test machine

  A。 formulated neutral salt spray test machine salt spray solution will analytically pure sodium chloride dissolved in purified water at a concentration of (50 ± 10) g / L。 Measured by PH paper test solution at 25 ℃, PH value of 6。5 to 7。5。 May be added when out of range of analytical grade hydrochloric acid or sodium hydroxide solution to adjust the preparation of a good solution before use subject to filtering。

  B. acetic acid salt spray test machine - first formulated analytically pure salt spray solution of sodium chloride dissolved in purified water at a concentration of (50 ± 10) g / L, and then the solution was adjusted with acetic acid PH value, so it is within the range of 3.1 to 3.3, the PH value of the collected samples with Salt measured. PH value should be measured at 25 ℃, the prepared solution was filtered before use required.

  C. formulated copper accelerated salt spray test machine acetic acid / salt spray solution

  The first step would be consistent with analytical solution of sodium chloride in compliance with GB6682XX water at a concentration of (50 ± 10) g / L.

  The second step was added in 3.8 liters of reagent grade copper chloride brine to dissolve thoroughly mixed.

山东11选5开奖结果   The third step, with acetic acid to adjust the PH value of the solution that fits within the 3.1 to 3.2 range, the PH value with